General Description
GLP1 and GLP2 type stations are electronically designed, programmable automatic high-voltage generators.
The devices are particularly suitable for manual use with clamps or test guns or for use in an automatic system.
Main Features
- High Voltage Test at 5.5kVAC max and 6kVDC
- Leakage Current up to 3mA AC and 10mA DC
- Insulation Test with max voltage 6kVDC and max resistance 10GΩ
- Digital display of measurements on control panel
- Programming and messaging display on control panel with keypad
- All programmable parameters (voltage, duration, minimum threshold, maximum threshold, ramp-up/ramp-down, etc.)
- In case of fault: circuit breaker trip, audible and visual alarm
- Test recording
- Numerous interfaces: USB, RS232, Ethernet, digital inputs/outputs, etc..
- PC control possible
Technical specifications
| Power Supply | 230 V / 50 Hz + ground |
| Presentation | standard 14″ box (also available in 19″ case BB1880B or table case BB1800B) |
| High-voltage AC/DC test
Voltage adjustment and measurement
Leakage current measurement Trip-out threshold Test duration Burning function
Insulation test Voltage adjustment and measurement
Insulation resistance measurement Fault threshold Test duration |
Adjustable up to 5.5KVAC and 6KVDC Digital Display Digital Display : 0 to 3mAAC and 0 to 10mADC Programmable minimum and maximum thresholds Programmable Without
Adjustable up to 6KVDC Digital Display Digital Display : 250kΩ to 10GΩ Programmable Programmable |
| Operation
Ground reference Test launch mode Recording results |
Floating potential Cycle start button or hold button on remote control input depending on settings With |
Items supplied with the unit :
- User manual
- Verification report with COFRAC connection certificate
Available accessories :
- Test cables
- Pistolets de test BB02 ou BB03
- Fault simulation box
- Test cages
- DLL library and driver software for PC
- …