General Description
GLP1 and GLP2 type stations are electronically designed, programmable automatic insulation resistance measurement stations.
The devices are particularly suitable for manual use with clamps or test guns or for use in an automatic system.
Main Features
- Four-wires method measurement
- High Voltage Test from 100V to 3000VDC
- Leakage current up to 4mA DC
- Insulation test with voltage from 100V to 3000VDC
- Resistance measurement from 500kΩ to 1.5GΩ
- Digital display of measurements on a touchscreen panel
- Programming and messaging display on the touchscreen panel
- All programmable parameters (voltage, duration, minimum threshold, maximum threshold, ramp-up/ramp-down, etc.)
- In case of fault: circuit breaker trip, audible and visual alarm
- Test recording
- Numerous interfaces: USB, RS232, Ethernet, digital inputs/outputs, etc..
- PC control possible
Technical specifications
| Power Supply | 230 V / 50 Hz + ground |
| Presentation | standard 14″ case |
| High-voltage AC/DC test
Voltage adjustment and measurement
Leakage current measurement Trip-out threshold Test duration Burning function
Insulation test Voltage adjustment and measurement
Insulation resistance measurement Fault threshold Test duration |
Adjustable from 100VDC to 3000VDC Digital Display Digital Display : 0 to 4mADC Programmable minimum and maximum thresholds Programmable Without
Adjustable from 100VDC to 3000VDC Digital Display Digital Display : 250kΩ to 1.5GΩ Programmable Programmable |
| Operation
Ground reference Test launch mode Recording results |
Floating potential Cycle start button or hold button on remote control input depending on settings Results recorded |
Items supplied with the unit :
- User manual
- Verification report with COFRAC connection certificate
Available accessories :
- Test cables
- Pistolets de test BB02 ou BB03
- Fault simulation box
- Test cages
- DLL library and driver software for PC
- …