General Description
The GLP1 and GLP2 types are programmable, electronically designed multi-test devices.
With their high-voltage leakage current limited to 4mA, they can be used without special safety devices (EN 50191 standard).
They allow for the sequencing of various safety tests required by different standards (Mass Continuity, Insulation, High Voltage in AC or DC) as well as functional tests (single-phase product powered at its nominal voltage by an external source).
All tests are automatically performed sequentially with a single product connection.
Main Features
- Continuity Test of Live Parts at 24VDC for detecting phase/neutral connections or short circuits
- Continuity Test of Ground at 6V or 12VAC – Current from 10A to 30A – Programmable resistance threshold – 4-wires method
- High Voltage Test at 5.5KVAC max and 6KVDC – Leakage current up to 3mA AC and 10mA DC
- Insulation Test with maximum voltage 6KVDC and maximum resistance 10GΩ
- Simulation Test of Standard Leakage (without isolation transformer and with summation of phase and neutral leakage)
- Functional Test with external voltage source for current consumption measurement
- Digital display of measurements on control panel
- Programming and messaging display on control panel with keypad
- All programmable parameters (voltage, duration, minimum threshold, maximum threshold, ramp-up/ramp-down, etc.)
- In the event of a fault : disconnection, audible and visual alarm
- Test recording
- Numerous interfaces : USB, RS232, Ethernet, Digital I/O…
- PC control possible
- Two presentation modes : tabletop unit — 19″ enclosure unit for integration
- Two types of control panels : keyboard panel — touchscreen panel
Technical specifications
| Power Supply | 230 V / 50 Hz + ground |
| Presentation | standard 14″ case (also available in 19″ case BB1880B or table case BB1800B) |
| Ground Continuity Test
Open-Load Voltage Setting Test Current Setting Resistance Measurement Test duration
Continuity Test of Live Parts Voltage Setting Current Measurement Fault Threshold Test duration
AC/DC High Voltage Test Voltage Setting and Measurement Leakage Current Measurement Trip Threshold Test duration
Insulation Test Voltage Setting and Measurement Insulation Resistance Measurement Fault Threshold Test duration
Standard Leakage Test (Simulation) Voltage Setting and Measurement
Leakage Current Measurement Fault Threshold Test duration
Functional Test Voltage Setting and Measurement
Current Consumption Measurement Fault Threshold Test duration Other Measurements |
6VAC or 12VAC
10A to 30A Digital display 0 to 400mΩ or 0 to 1100mΩ depending on the version Configurable
Fixed to 24VDC Digital display 0 to 600mA Configurable minimum and maximum values Fixed < 1s
Adjustable up to 5.5KVAC and 6KVDC Digital Display: 0 to 3mAAC and 0 to 10mADC Programmable minimum and maximum thresholds Programmable
Adjustable up to 6KVDC Digital display Digital display : 250kΩ to 10GΩ Programmable Programmable
Adjustable from 10VAC to 270AC Digital display Digital display : 0 to 10mA Programmable Programmable
External source or (internal limited to 1A or 2A depending on version) With or without voltage measurement – Digital display Digital Display : 0 to 16A – Automatic range change depending on version Programmable Programmable Optional features include power, cosPhi, resistance measurements… |
| Operation
Ground reference Test Launch Mode Results Recording |
Floating potential Cycle start button or hold button on remote control input depending on settings With |
Items supplied with the unit :
- Instruction manual
- Verification report with COFRAC-certified connection proof
Available accessories :
- Ground continuity probe
- Control box (two-hand) and connection box
- Fault simulation box
- Test cages
- DLL library and PC control software
- …