Description générale
GLP1 and GLP2 type stations are leakage current measurement stations of electromechanical design.
The devices are particularly suitable for manual use with clamps or test guns or for use in an automatic system.
Main Features
- Measurement of contact current and current in the protective conductor (EN60990 standard)
- Voltage source : 0 to 250VAC
- Leakage current measurement : 1mA and 10mA ranges
- Maximum current consumption : 5A – Isolated source
- Analog display of voltage/current measurements: Class 1.5
- Programmable parameters via front panel buttons
- In case of fault : circuit breaker tripping, audible and visual alarm
Technical specifications
| Power Supply | 230 V / 50 Hz + ground |
| Presentation | standard 19″ box |
| Leakage current testing
Voltage adjustment and measurement Leakage current measurement Fault threshold |
Adjustable from 0 to 250VAC – Class 1.5 analog measurement 2 ranges of class 1.5 analog measurement: 1mA and 10mA Adjustable via rotary knob |
| Operation
Ground reference Test Launch Mode Results Recording |
No Cycle start button Without |
Items supplied with the unit :
- User manual
- Verification report with COFRAC connection certificate
Available accessories :
- Test cages
- DLL library and driver software for PC
- …